Status Mentor PGD2 Detector: Do Not Use Until…

 Safety Alerts  Comments Off on Status Mentor PGD2 Detector: Do Not Use Until…
Aug 252011
 

Status Mentor PGD2 gas meter: Do not use without the correct software

If you have a Status Mentor PGD2 gas meter do not use it until you have verified that the correct software version is installed and that the low battery alarm point is correctly configured in accordance with the instructions provided on the Status Scientific Controls website: www.status-scientific.com/software-downloads.cfm , warns the UK Health & Safety Executive.

Tests by the Health & Safety Laboratory (HSL) on behalf of HSE and discussions with Status Scientific, the manufacturer, have determined that the low battery voltage cut-off setting may be incorrect in some instruments having an older version of software.

This older software may allow the instrument to continue to remain active when there is insufficient power for it to work accurately. In this case the indicated measurement may be below the actual gas concentration in the sampled atmosphere and the alarm may not be activated when the actual gas concentration reaches the alarm set-point.

This Safety Notice is being issued as a result of investigations by the Health and Safety Executive into the Status Scientific Controls portable gas detector type Mentor PGD2. This instrument is intended to be used for the protection of personnel entering or working in an environment where there is a possibility of flammable or toxic gas being present. Continue reading »